A-96.250.441 / 060619
AMU pH-Redox
Installation
3.9.
Power supply
CAUTION
Do not apply power to the transmitter until all electrical connec-
tions have been made.
Installation
requirements
The installation must meet the following requirements:
Mains cable according to standards IEC 60227 or IEC
60245; flammability rating FV1
Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMU pH-Redox
3.10. Sensor
Connect the pH or ORP sensor to the AMU transmitter according to
the connection diagram, see
For sensor settings, see
3.11. Flow Meter
Connect the flow meter (if any) to the AMU transmitter according to
the connection diagram, see
3.12. Input
NOTICE:
Use only potential-free (dry) contacts.
Terminals 10/11
For programming see
Program List and Explanations, p. 61
.
A
B
C
Power supply connector
Phase conductor
Neutral conductor
A
B
C