AMI Solicon4
Error List
Error
Description
Corrective action
E001
Cond. Alarm high
– check process
– check programmed value, see
E002
Cond. Alarm low
– check process
– check programmed value, see
E003
Conc. Alarm high
– check process
– check programmed value, see
E004
Conc. Alarm low
– check process
– check programmed value, see
E007
Sample Temp. high
– check process
– check programmed value, see
E008
Sample Temp. low
– check process
– check programmed value, see
E009
Sample Flow high
– check sample inlet pressure
– check programmed value, see
E010
Sample Flow low
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
E011
Temp. shorted
– Check wiring of sensor
E012
Temp. disconnected
– Check wiring of sensor
E013
Case Temp. high
– check case/environment temperature
– check programmed value, see
E014
Case Temp. low
– check case/environment temperature
– check programmed value, see