AMI Solicon4
Program List and Explanations
A-96.250.321 / 191015
61
Control upwards or downwards
Setpoint:
User-defined process value for the selected parameter.
P-Band:
Range below (upwards control) or above (downwards con-
trol) the set-point, within the dosing intensity is reduced from 100%
to 0% to reach the setpoint without overshooting.
5.2.1.43
Control Parameters:
if Parameters = Conductivity
5.2.1.43.10
Setpoint
Range: 0
S–300 mS
5.2.1.43.20
P-Band:
Range: 0
S–300 mS
5.2.1.43
Control Parameters:
if Parameters = Temperature
5.2.1.43.11
Setpoint
Range: -25 to +270 °C
5.2.1.43.21
P-Band:
Range: -25 to +270 °C
5.2.1.43
Control Parameters:
if Parameters = Sample flow
5.2.1.43.12
Setpoint
Range: 0 –50 l/h
5.2.1.43.22
P-Band:
Range: 0 –50 l/h
5.2.1.43
Control Parameters:
if Parameters = Cond. uc.
5.2.1.43.13
Setpoint
Range: 0
S–300 mS
5.2.1.43.23
P-Band:
Range: 0
S–300 mS
5.2.1.43.3
Reset time:
The reset time is the time till the step response of a sin-
gle I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9’000 Sec
5.2.1.43.4
Derivative time:
The derivative time is the time till the ramp re-
sponse of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0–9’000 Sec
5.2.1.43.5
Control timeout:
If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped for safety reasons.
Range: 0–720 min