AMI Solicon4
Product Description
A-96.250.321 / 191015
7
2.
Product Description
2.1.
Description of the System
This instrument is applicable for the measurement of the specific
conductivity in surface water, potable water and cooling water. The
complete system consists of the Transmitter AMI Solicon4, the con-
ductivity sensor Swansensor Shurecon P and the flow cell
M-Flow PG.
Transmitter, flow cell and sensors are also available as single com-
ponents.
Application
Range
Conductivity is a parameter for the total quantity of ions present in
the solution. It can be used for the controlling of:
the condition of waters
water purification
water hardness
completeness of ion analysis
Signal
Outputs
Two signal outputs programmable for measured values (freely scal-
able, linear, bilinear, log) or as continuous control output (control
parameters programmable).
Current loop:
0/4–20 mA
Maximal burden:
510 Ohm
Third signal output with the same specifications as option.
(Only possible if no communication interface is used.)
Relays
Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A/250 VAC
Alarm Relay
One potential free contact.
Alternatively:
Open during normal operation, closed on error or loss of
power.
Closed during normal operation, open on error or loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.