Maintenance
A-96.250.321 / 191015
41
6.3.
Maintenance of the Sensor
6.3.1
Remove the Sensor form the Flow Cell
To remove the sensor form the flow cell proceed as follows:
1
Unscrew and remove the fixing sleeve [A].
2
Pull the conductivity sensor [B] out of the flow cell block [B].
Cleaning
If the sensor is contaminated, take a small brush and clean it with
water and detergents.
In case of heavy contamination with oil or grease, use ethanol to
clean it. Take a soft tissue and clean the tip of the sensor cautious-
ly.
After each cleaning, the sensor must be rinsed with clean water.
6.3.2
Install the Sensor into the Flow Cell
1
Make sure that the washer [C] and the O-ring [D] are in correct
position.
2
Push the sensor through the flow cell bock [E] into the flow cell.
3
Tighten the fixing sleeve [A] to fix the sensor.
A
B
C
D
E
Fixing sleeve
Conductivity sensor
Teflon washer
O-Ring
Flow cell block
A
C
D
B
E