AMI Solicon4
Program List and Explanations
5.3.1.5
Alarm Concentration:
Visible if a Conc. parameter has been se-
lected. TDS is displayed in mg/l, all other parameters are displayed
in %.
5.3.1.5.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003, is displayed in the mes-
sage list.
Range: 0.00%–99.90%
5.3.1.5.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00%–99.90%
5.3.1.5.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range.0.00%–99.90%
5.3.1.5.45
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above or fallen below the pro-
grammed alarm.
Range: 0–28‘800 Sec
5.3.2 and 5.3.3
Relay 1 and 2:
The contacts can be set as normally open or nor-
mally closed with a jumper. See
. The
function of relay contacts 1 or 2 are defined by the user.
Note:
The navigation in the menu <Relay 1> and <Relay 2> is
identical. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1
First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2
Then enter the necessary data depending on the selected func-
tion.