AMI Solicon4
Program List and Explanations
A-96.250.321 / 191015
63
5.3.1.2
Sample Flow:
Define at which sample flow an alarm should be is-
sued.
5.3.1.2.1
Flow Alarm:
Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
Note:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.2.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 9–20 l/h
5.3.1.2.35
Alarm Low:
If the measuring values falls below the programmed
value E010 will be issued.
Range: 5–8 l/h
5.3.1.3
Sample Temp.
5.3.1.3.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E007, is displayed in the mes-
sage list.
Range: 30–200 °C
5.3.1.3.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C
5.3.1.4
Case Temp.
5.3.1.4.1
Case Temp. high:
Set the alarm high value for temperature of elec-
tronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.4.2
Case Temp.
low:
Set the alarm low value for temperature of elec-
tronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to +20 °C