A-96.250.451 / 190121
AMI Deltacon Power
Program List and Explanations
63
5.1.2.2.2
Temp. Corr
: Enter the temperature correction printed on the sensor
label.
5.1.2.2.3
Cable length
: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2.2.5
Temp. comp
:
5.1.2.2.5.1
Comp
.: Available compensation models:
Strong acids
5.1.3
Flow
5.1.3.1
Sensor
: Choose between <Yes> od <No>. <Yes> is set by default. If
Sensor is set to <No>, the flow monitoring is deactivated and the sig
-
nal <Sample flow> on the Signal Outputs and Relay 1 and 2 is not
available.
5.2 Signal Outputs
Note:
The navigation in the menu <Signal Output 1> and <Signal
Output 2> is equal. For reason of simplicity only the menu
numbers of Signal Output 1 are used in the following.
5.2.1
Signal Output 1:
Assign process value, the current loop range and a
function to each signal output.
5.2.1.1
Parameter:
Assign one of the process values to the signal output.
Available values:
Cond 1 (cc)
Cond 2 (sc)
Temp. 1
Temp. 2
Difference
Sample flow
pH
Ammonia
5.2.1.2
Current Loop:
Select the current range of the signal output.
Make sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3
Function:
Define if the signal output is used to transmit a process val
-
ue or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
Control upwards or control downwards for controllers.
See