A-96.250.451 / 190121
AMI Deltacon Power
Safety Instructions
9
1.3. Restriction for use
The AMI Deltacon Power is designed for determination of:
specific (total) conductivity
cation (acid) conductivity after the cation exchanger
in power plant water.
It calculates the pH value and the concentration of the alkaline sub
-
stance (NH
3
, morpholine, etc.) if an alkaline substance is present in
the water.
It is not suitable for pH determination in high purity water before
alkalization agent addition.
Conditions for pH calculation:
only 1 alkalization agent in the sample
the contamination is mostly NaCl
phosphate concentration is < 0.5 ppm
pH value is > 7.5, and < 11.5
if pH value is < 8, the concentration of contaminant must be
small compared to the concentration of the alkalization agent
No sand. No oil.
The sample must not contain any particles, which may block the flow
cell. Sufficient sample flow is coercive for the correct function of the
instrument.
WARNING
For safe instrument installation and operation you must read and
understand the instructions in this manual, as well as the Material
Safety Data Sheets (MSDS)
Cation Exchange Resin
Download
MSDS
The current Material Safety Data Sheets (MSDS) for the below listed
Reagents are available for downloading at
www.swan.ch
.
Product name:
Cation Exchange Resin
Catalogue number:
A-82.841.030, A-82.841.031 and
A-82.841.035