Troubleshooting
50
E015
pH Calculation undef.
– Calculated pH value < 7.5 or > 11.5
E017
Control time-out
– Check control device or programming
in Installation, Relay contact, Relay 1/2
E019
Temp. 2 shortened
– check wiring of temperature sensor
– check temp. sensor
E020
Temp. 2 disconnected
– check wiring of temperature sensor
– check temp. sensor
E024
Input active
– See If Fault Yes is programmed in
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2
E030
EEProm Frontend
– call service
E031
Cal. Recout
– call service
E032
Wrong Frontend
– call service
E033
pH Alarm high
– check process
– check programmed value, see
E034
pH Alarm low
– check process
– check programmed value, see
E035
Alkali Alarm high
– check process
– check programmed value, see
E036
Alkali Alarm low
– check process
– check programmed value, see
E037
Temp. 2 Alarm high
– check process
– check programmed value, see
Error
Description
Corrective action