A-96.250.451 / 190121
AMI Deltacon Power
Program List and Explanations
60
Press the <Enter> key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val
-
ues will be reset.
3.2 Set Time
Adjust date and time.
3.3 Change of Resin
If <Monitoring of Resin> in menu <Installation>/<Miscellaneous>
is set to <Yes>, the parameter <Change of Resin> has to be set to
<Yes> after each time the cation exchanger bottle has been re
-
placed.
4 Operation
4.1 Sensors
4.1.1
Filter Time Constant:
Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.1.2
Hold after Cal.:
Delay permitting the instrument to stabilize again af
-
ter calibration. During calibration plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac
-
tive.
Range: 0–6‘000 Sec
4.2 Relay Contacts
See
Alarm relay:
Active or inactive
Relay 1/2:
Active or inactive
Signal output 1/2:
Actual current in mA
Signal output 3:
Actual current in mA (if option is installed)