AMI Deltacon Power
Program List and Explanations
59
2.2.2
Miscellaneous
:
2.2.2.1
Case Temp:
Shows the current temperature in [°C] inside the trans
-
mitter.
2.3 Sample
2.3.1
Sample ID
:
Shows the identification assigned to a sample. This iden
-
tification is defined by the user to identify the location of the sample.
2.3.2
Sample Flow
: Shows the current sample flow in l/h and the Raw
Value in Hz.
The Sample flow must be above 5 l/h.
2.3.3
Sample Temp:
2.3.3.1
Temp 1:
Shows the current sample temperature at sensor 1 in °C.
(Pt 1000)
: Shows the current temperature at sensor 1 in Ohm.
Temp 2:
Shows the current temperature at sensor 2 in °C.
(Pt 1000)
: Shows the current temperature at sensor 2 in Ohm.
2.4 I/O State
Shows current status of all in- and outputs.
2.4.1/2.4.2
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
3 Maintenance
3.1 Simulation
To simulate a value or a relay state, select the
alarm relay
relay 1/2
signal output 1/2
with the [
] or [
] key.
Press the <Enter> key.
Change the value or state of the selected item with the [
] or
[
] key.
Alarm relay:
Active or inactive.
Relay 1/2:
Active or inactive.
Input:
Open or closed.
Signal output 1/2:
Actual current in mA
Signal output 3:
Actual current in mA (if option is installed)