AMI INSPECTOR Hydrogen
Default Values
A-96.250.741 / 050218
67
10. Default Values
Operation:
Sensors:
Filter Time Const.: ..................................................................... 10 s
Hold after Cal.:......................................................................... 300 s
Alarm Relay
...................................................................... same as in Installation
Relay 1/2
...................................................................... same as in Installation
Input
...................................................................... same as in Installation
Logger:
Logger Interval:..................................................................... 30 min
Clear Logger:................................................................................ no
Installation:
Sensors
Miscellaneous; Flow: ............................................................... None
Miscellaneous; Offset: .......................................................... 0.0 ppb
Miscellaneous; Maintenance Int. ................................................. 3 h
Quality Assurance; Level: ........................................................ 0: Off
Sensor parameters; Saturation current: ....................................... ---
Sensor parameters; Air pressure:............................................0 hPa
Signal Output
Parameter:........................................................................ Hydrogen
Current loop:..................................................................... 4 –20 mA
Function:.................................................................................. linear
Scaling: Range low:............................................................ 0.00 ppb
Scaling: Range high: ....................................................... 10.00 ppm
Alarm Relay:
Alarm Hydrogen; Alarm high:........................................... 10.00 ppm
Alarm Hydrogen; Alarm low: ............................................... 0.00 ppb
Alarm Hydrogen; Hysteresis:............................................... 100 ppb
Alarm Hydrogen; Delay: ............................................................ 30 s
If Flow = Q-Flow
Sample Flow, Flow Alarm: .......................................................... yes
Sample Flow, Alarm high: .....................................................14.0 l/h
Sample Flow, Alarm low: ........................................................6.0 l/h
Sample Temp., Alarm High: ..................................................... 50 °C
Sample Temp., Alarm Low:........................................................ 0 °C
Alarm Saturation; Alarm high..................................................120 %
Alarm Saturation; Alarm low ....................................................0.0 %
Alarm Saturation; Hysteresis ......................................................2 %
Alarm Saturation; Delay............................................................. 30 s
Case temp. high: ..................................................................... 65 °C
Case temp. low:......................................................................... 0 °C