AMI INSPECTOR Hydrogen
Maintenance
A-96.250.741 / 050218
37
6.4.5
Completion of the measurement
1
Stop the sample flow to the AMI INSPECTOR Hydrogen by
closing the appropriate valve, e.g. back pressure regulator,
sample preparation or flow regulating valve at flow cell again.
2
Close flow regulating valve of the AMI Inspector.
3
Disconnect the AMI Inspector by removing the tubes and con-
nect the sample outlet of the Monitor AMI INSPECTOR Hydro-
gen to the sample outlet funnel again.
4
Start sample flow again and regulate sample flow.
5
Shutdown the AMI INSPECTOR Hydrogen.
If the instrument will not be used for a longer period of time, see
Longer Stop of Operation, p. 39
.