12
A-96.250.741 / 050218
AMI INSPECTOR Hydrogen
Product Description
2.3.
Technical Data
Power Supply
Battery
Use original, supplied power adapter only.
Voltage:
Power consumption:
Charging time:
Battery type:
85–265 VAC, 50/60 Hz
max. 20 VA
6h
Li-Ion
During charging protect from heat impact and keep splash-proof
(not IP66).
Operating time
Stand-alone (Battery):
Connected adapter:
> 24h
continuous
Controlled shut-down when battery is empty, remaining time is dis-
played.
Electronics
Aluminium with a protection degree of IP 66 / NEMA 4X
housing
Ambient temperature:
Humidity:
Display:
-10 to +50 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
6 to 20 l/h
up to 45 °C
0.2 to 1 bar
pressure free
Flow cell and
connection
Flow cell made of acrylic glass with built-in flow adjustment valve
and digital sample flow meter
Sample inlet:
Sample outlet:
1/4” Swagelok tube adapter
flexible tube 8x6 mm
Measuring
Range
Range
0.1 to 199 ppb
200 to 800 ppb
0 – 50% saturation
Resolution
0.1 ppb
1 ppb
0.1% Saturation
Accuracy
Reprodu-
cibility
±3 % of measured value or ± 0.5 ppb
±1 % of measured value or ± 0.5 ppb