68
A-96.250.741 / 050218
AMI INSPECTOR Hydrogen
Default Values
Relay 1
Function: .......................................................................... limit upper
Parameter: ........................................................................ Hydrogen
Setpoint:...........................................................................10.00 ppm
Hysteresis: ...........................................................................100 ppb
Delay:......................................................................................... 30 s
Relay 2
Function: .......................................................................... limit upper
Parameter: ....................................................................Temperature
Setpoint:................................................................................50.0 °C
Hysteresis: ..............................................................................1.0 °C
Delay:......................................................................................... 30 s
If Function = Control upw. or dnw:
Parameter: .................................................................... Meas. Value
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................. 120/min
Settings: Control Parameters: Setpoint: .......................10.00 ppm
Settings: Control Parameters: P-band:.............................100 ppb
Settings: Control Parameters: Reset time: ............................... 0 s
Settings: Control Parameters: Derivative Time: ....................... 0 s
Settings: Control Parameters: Control Timeout:................... 0 min
Settings: Act. Time prop.: Cycle time: .................................... 60 s
Settings: Act. Time prop.: Response time: ............................ 10 s
Settings: Act. Motor valve: Run time: .................................... 60 s
Settings: Act. Motor valve: Neutral zone: ................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval:................................................................................. 1 min
Mode: ....................................................................................... daily
Start time: ........................................................................00.00.00
Mode:.....................................................................................weekly
Calendar; Start time: .......................................................00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ................................................................................... 10 s
Delay:........................................................................................... 5 s
Signal output:............................................................................. cont
Output/Control: .......................................................................... cont
Input:
Active ............................................................................when closed
Signal Outputs ........................................................................... hold
Output/Control .............................................................................. off
Fault..............................................................................................no
Delay.......................................................................................... 10 s