AMI INSPECTOR Hydrogen
Maintenance
A-96.250.741 / 050218
31
6.3.3
Replace Hydrogen Sensor
To replace the hydrogen sensor proceed as follows:
Remove the
old sensor
1
Switch off the AMI Inspector Hydrogen.
2
Close the flow regulating valve [D].
3
Unscrew and remove the threaded sleeve [A] from the flow cell.
4
Remove the hydrogen sensor [C] from the flow cell.
5
Unscrew and remove the sensor connector [B] from the hydro-
gen sensor.
Install the new
sensor
1
Screw the connector onto the hydrogen sensor and tighten it.
2
Put the hydrogen sensor into the flow cell.
3
Screw the threaded sleeve into the thread of the flow cell to fix
the hydrogen sensor.
4
Open the flow regulating valve and adjust the sample flow to
6–20 l/h.
5
Switch on the AMI Inspector Hydrogen.
6
Let the new sensor run in for at least 1h.