A-96.250.881 / 070622
AMI Codes-II O3
Program List and Explanations
73
5 Installation
5.1 Sensors
5.1.1
Disinf:
The measured value can be displayed in two ways:
Ozone: the measured value is displayed as ppb O
3
TC1: the measured value is displayed as ppb Cl
2
5.1.2
Reag. expiry warning:
If activated, message E066 will be displayed
as soon as the reagents are older than 30 days.
Choose between <Yes> or <No>.
5.2 Signal Outputs
5.2.1 and 5.2.2
Signal Output 1 and 2:
Assign process value, the current loop
range and a function to each signal output.
Note:
The navigation in the menu <Signal Output 1> and <Signal
Output 2> is equal. For reason of simplicity only the menu
numbers of Signal Output 1 are used in the following.
5.2.1.1
Parameter:
Assign one of the process values to the signal output.
Available values:
DIS
Sample flow
5.2.1.2
Current Loop:
Select the current range of the signal output.
Make sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3
Function:
Define if the signal output is used to transmit a process val
-
ue or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
Control upwards or control downwards for controllers.
See