A-96.250.881 / 070622
AMI Codes-II O3
Instrument Setup
35
4.4. Fill or Flush Reagent System
Fill or flush the reagent tubing:
upon the initial instrument setup,
after refilling the reagent canisters,
before a system shut-down to flush the system with
demineralized water until no more reagent is left in the
system.
1
Check tubing and flow cell for leaks and repair if necessary.
2
Let the instrument run continuously for 1 hour.
4.5. Programming
Programming
Program all parameters for external devices (interface, recorders,
etc.).
Program all parameters for instrument operation (disinfectant, limits,
alarms).
Program the reference values of the Verikit in menu <Operation/
Sensors/Verification kits>.
See
Program List and Explanations, p. 69
Navigate to menu< Maintenance /
Service/Fill system>.
Press [Enter].
The peristaltic pump is activated for
1.5 minutes.
Press [Exit] 4 x to go back to the oper
-
ating mode.