A-96.250.881 / 070622
AMI Codes-II O3
Installation
26
3.5.3
Opening the peristaltic pump housing
For some electrical connections (see
), it
is necessary to open the housing of the peristaltic pump. To do this,
proceed as follows:
1
Switch off the analyzer according to
2
Remove the protection cap and all pump tubes as described in
.
3
Unscrew the 4 screws of the peristaltic pump housing and re
-
move the cover.
4
Disconnect the motor connector [A].
5
Feed the cable into the housing through one of the PG7 cable
glands (permissible cable thicknesses are specified in
).
6
Connect the cable to the terminal block of the peristaltic pump ac
-
7
Reassemble in reverse order.