![Swan Analytical Instruments AMI Codes-II O3 Operator'S Manual Download Page 69](http://html1.mh-extra.com/html/swan-analytical-instruments/ami-codes-ii-o3/ami-codes-ii-o3_operators-manual_1411369069.webp)
A-96.250.881 / 070622
AMI Codes-II O3
Program List and Explanations
69
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active,
acknowledged). If an active error is acknowledged, the alarm relay is
active again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5
Indicates necessary maintenance, e.g. preparing new reagents.
1.3 Message List
1.3.1
Shows the error history: Error code, date and time of issue and
status (active, acknowledged, cleared). 65 errors are memorized.
Then the oldest error is cleared to save the newest error (circular
buffer).
1.4 Audit Trail
Shows the audit trail: event, menu, date and time of issue. 96 events
are memorized. Then the oldest event is cleared to save the newest
one (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation
:
View the designation of the instrument.
Version
: Firmware of instrument (e.g. V6.21-01/18)
2.1.3
Peripherals
: PeriClip 1: Firmware of peristaltic pump (e.g. 1.05)
2.1.4
Factory Test
: Test date of the instrument, motherboard and frontend
QC factory test.
2.1.5
Operating Time
: Years/days/hours/minutes/seconds
2.2 Sensors
2.2.1
Photometer:
Current value
: Shows the actual photometer signal [ppb].
Raw value
: Shows the actual photometer signal [Hz].
Absorbance