A-96.250.881 / 070622
AMI Codes-II O3
Program List and Explanations
71
3 Maintenance
3.1 Calibration
3.1.1
Process DIS
: Possibility to correct the disinfectant value.
See
Process Calibration of DIS, p. 50
, for more details.
3.2 Service
3.2.1
Verification: Performs a verification using the reference kit. Follow
dialog. See
.
3.2.2
Fill System: Activates the reagent pump.
3.2.3
Reagents:
Displays the date of the last filling and the remaining
number of days until the maximum storage time of 30 days is
exceeded.
3.2.3.3
Change reagents
: Resets the counter and saves the date of the new
filling.
3.3 Simulation
To simulate a value or a relay state, select the
alarm relay,
relay 1 or 2
signal output 1 or 2
with the [
] or [
] key.
Press the [Enter] key.
Change the value or state of the selected item with the [
] or
[
] key.
Press the [Enter] key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val
-
ues will be reset.
3.4 Set Time
Adjust date and time.
Alarm Relay:
Active or inactive
Relay 1 and 2:
Active or inactive
Signal Output 1 and 2:
Actual current in mA
Signal Output 3:
Actual current in mA (if option is installed)