Maintenance
42
6.2. Stop of Operation for Maintenance
1
Put the suction lances into a bucket with demineralized water.
2
Start <Fill system>.
The reagent tubes are flushed with water.
3
Remove the suction lances from the water.
4
Start <Fill system> again.
The water will be pumped out of the reagent tubes.
5
Stop sample flow.
6
Wait until level in flow cell is empty.
7
Shut off power of the instrument.