
28
[Fig. IV-2-4] SLM Integration Time Setting
[Fig. IV-2-6] Measured Results
2. 3 Detector
The results of 3 detectors are displayed in vibration level meter and 2 detectors are displayed in
sound level meter simultaneously. Only the data of the selected detector are displayed and stored. For
example, if you want to measure acceleration in vibration level meter, you can select only a detector
like [Fig. IV-2-7]. And if you want to measure Leq in sound level meter, you can select only a
detector like [Fig
.
IV-2-9]. And [Fig. VI-2-8], [Fig. VI-2-10] is shown the activated display window.
The options of each detector are listed vertically as shown in [Fig. VI-2-7] and [Fig. VI-2-9].
RealWave supports various units with gravity acceleration(g) and decibel(dB) (see [Fig. IV-2-11] and
[Fig. IV-2-12]).
[Fig. IV-2-7] One Detector Selection in VLM
[Fig. IV-2-8] Activated Window for One
Detector in VLM
Summary of Contents for SLA-PA201
Page 64: ...64 Appendix 1 The Theory for calculating RMS by ISO2631 and ISO8041 ...
Page 67: ...67 Appendix 2 Frequency Response of Filter in the Vibration Level Meter ...
Page 75: ...75 Appendix 3 Theory of WBCombined Filter and Frequency Response ISO6954 Filter ...
Page 79: ...79 Fig 2 Lowpass Filter with 100Hz Cutoff Fig 3 a v transition Frequency Response ...
Page 88: ...88 Fig 12 Error between theoretical value and measured value ...