background image

User

 

Manual

U-

SUP-DC2000-EN1

Universal controller

Supmea Automation Co.,Ltd

.

www.supmea.com

Headquarters

5th floor,Building 4,Singapore Hangzhou Science Technology Park,No. 6 street,

 

Hangzhou Economic Development Area,Hangzhou 310018,China

Singapore 

2 Venture Drive #11-30 Vision Exchange Singapore

[email protected]

Summary of Contents for SUP-DC2000

Page 1: ...ea Automation Co Ltd www supmea com Headquarters 5th floor Building 4 Singapore Hangzhou Science Technology Park No 6 street Hangzhou Economic Development Area Hangzhou 310018 China Singapore 2 Venture Drive 11 30 Vision Exchange Singapore info supmea com ...

Page 2: ...to incorrect operation After you finish reading please keep it in a place where it can be easily accessed at any time for reference during operation Note The contents of this manual are subject to change without notice due to functional upgrades We strive to be correct in the contents of this manual If you find any errors please contact us The contents of this manual are strictly prohibited to be ...

Page 3: ...cedures may impair the protection provided by this product The company does not assume any responsibility for the quality performance function and product safety issues arising from the above conditions When installing lightning protection devices for this product and its control system or designing and installing separate safety protection circuits for this product and its control system it needs...

Page 4: ...ork shall be grounded at is electric level shielded wires shall be located rationally SPD surge protector shall be applied properly Some inner parts may carry high voltage Do not open the square panel in the front except our company personnel or maintenance personnel acknowledged by our company to avoid electric shock Cut off electric powers before making any checks to avoid electric shock Check t...

Page 5: ... package Prevent dust wire end iron fines or other objects from entering the instrument during installation otherwise it will cause abnormal movement or failure During operation to modify configuration signal output startup stop operation safety shall be fully considered Operation mistakes may lead to failure and even destruction of the instrument and controlled equipment Each part of the instrume...

Page 6: ...d directly or indirectly by improper operation of the user Package contents No Name Quantity Remark 1 Controller 1 2 Butterfly buckle 2 For fixed installation 3 Manual 1 4 Certificate 1 5 Standard Accessories Package 1 After opening the box please confirm the package contents before starting the operation If you find that the model and quantity are incorrect or there is physical damage in appearan...

Page 7: ...nsor search interface 10 6 2 Sensor type selection interface 10 6 3 Monitoring interface 11 6 4 Password verification interface 12 6 5 Main menu interface 13 Chapter 7 Configuration Settings 14 7 1 System setting interface 14 7 2 Signal setting interface 14 7 3 Remote transmission setting interface 24 7 4 Alarm setting interface 25 7 5 Information query interface 27 Chapter 8 Communication Protoco...

Page 8: ...suitable for use with various water quality series digital sensors of our company It is used to monitor water quality parameters including pH ORP conductivity dissolved oxygen turbidity sludge concentration and other water quality parameters Through RS485 or current transmission output to the monitoring room for record keeping ...

Page 9: ...y Dissolved oxygen Turbidity MLSS Isolated transmission output is adopted which is less affected by interference Using isolated RS485 communication technology With high and low alarm output function With sound and light alarm function With LCD backlight switch control function improve the adaptability of ambient light ...

Page 10: ...ity 0 600 mS cm Turbidity 0 4000 NTU MLSS 0 120000 mg L Current output Isolation type 4 20 mA can set the corresponding parameter measurement range load capacity 500Ω output accuracy 0 2 FS RS485 output Isolated Modbus RTU communication protocol Alarm Normally open and normally closed 2 channels capacity AC250V 3A Distribution output 12V 125mA Relative humidity 10 85 no condensation Operating temp...

Page 11: ... Please do not tilt left and right when installing this product try to install it horizontally can be tilted back 30 Avoid the following places when installing Places directly exposed to sunlight and near hot appliances Places where the ambient temperature exceeds 60 C during operation Places where the ambient humidity exceeds 85 during work Near the source of electromagnetic generation Places wit...

Page 12: ...Chapter 4 Fixed installation www supmea com 5 Figure 3 Schematic diagram of meter installation Front panel Butterfly buckle ...

Page 13: ...Chapter 4 Fixed installation www supmea com 6 4 2 Instrument wiring Figure 4 220VAC wiring diagram ...

Page 14: ...Chapter 4 Fixed installation www supmea com 7 Figure 5 24VDC wiring diagram ...

Page 15: ...l negative HO High alarm normally open relay HC High alarm normally closed relay COM relay common terminal LO Low alarm normally open relay LC Low alarm normally closed relay COM relay common terminal L Power port L NC empty N Power port N 24V Positive pole of 24V power supply 24V 24V power supply negative Notice To prevent electric shock please confirm that the meter is not powered on before conn...

Page 16: ...itoring Interface Return to the upper layer between the relevant upper and lower layers of the interface under the menu interface Right shift Modified bits of the cycle selection parameter Menu Enter the menu under Monitoring Interface Exit menu under Menu interface Downshift Select the relevant menu under Menu interface Modify the relevant value in the setting state Enter Enter the submenu under ...

Page 17: ...toring interface If the match is not successful enter the sensor type selection interface Figure 7 6 2 Sensor type selection interface pH ORP Optional digital sensor model pH 8001 Turbidity optional digital sensor model PTU 8010 PTU 8011 PSS optional digital sensor model PSS 9011 Dissolved Oxygen Optional digital sensor model DO 7010 DO 7012 DO 7013 DO 7016 Conductivity optional digital sensor mod...

Page 18: ...ection are correct If the sensor type is incorrectly selected long press to re enter the sensor type selection interface Figure 9 Use to enter the password verification interface enter the password to enter the main menu interface Use to enter the alarm query interface to query the current alarm setting information Note XX automatically switches the measurement parameter type according to the sele...

Page 19: ... Note If the instrument is not operated for 3 minutes it will automatically jump to the monitoring interface 6 4 Password verification interface After entering the password use it to enter the main menu interface The initial password is 0000 you can use the password modification function to modify the password If you forget your password please contact our company Figure 16 DO 25 0 20 00mA 0 05mg ...

Page 20: ...gnal Settings Sensor parameters Calibration Revision and other function settings Remote setting Communication Settings Electrorheological send Alarm setting parameter setting of high alarm and low alarm Information inquiry current version information query Figure 17 Main Menu 1 Systeming Setting 2 Signal Setting 3 Remote Setting 4 Alarm Setting 5 Information Inquiry ...

Page 21: ...ing 7 2 1 1 Online calibration Wash the sensor with distilled water wipe off the water stains and then put it in the calibration solution Select the calibration point according to the calibration solution The corresponding values of the NIST calibration points from small to large are 4 00 pH 6 86 pH and 9 18 pH respectively the corresponding values from small to large in USA are 4 01 pH 7 00 pH an...

Page 22: ...n be corrected Figure 20 7 2 1 3 Measurement parameters Select the sensor to measure the primary parameter pH or ORP Figure 21 Signal Setting 1 Online Calibration 2 Signal Correction 3 Measure Param Signal Correction Signal Setting 1 Online Calibration 2 Signal Correction 3 Measure Param Measure Param 1 pH 2 ORP ...

Page 23: ...n wait for a while and use the to complete the calibration after the measured value is stable Before calibration the sensor needs to be washed with distilled water and wiped dry After soaking in water put it in the calibration solution for calibration Figure 22 7 2 2 2 Signal correction The measured ORP value can be corrected Figure 23 Signal Setting 1 Online Calibration 2 Signal Correction 3 Meas...

Page 24: ...The measured turbidity value can be corrected Figure 25 7 2 3 3 Wiping Time Set the sensor wipe time interval 1min 5min 15min 30min 1h 4h 12h 1d 3d 7d Figure 26 Signal Setting 1 Tu Factor 2 Signal Correction 3 Wiping Time Tu Factor Signal Setting 1 Tu Factor 2 Signal Correction 3 Wiping Time Signal Correction Signal Setting 1 Tu Factor 2 Signal Correction 3 Wiping Time Wiping Time ...

Page 25: ...e first point calibration interface of sludge concentration and wipe the sensor probe Put the probe into the first point of standard solution usually pure water is used as the first point input the standard value of the first point of standard solution and click after the measured value data is stable click to complete the first point calibration Take the probe out of the first standard solution r...

Page 26: ... setting of sludge concentration factor Figure 29 7 2 4 4 Wiping Time Set the sensor wipe time interval 1min 5min 15min 30min 1h 4h 12h 1d 3d 7d Figure 30 Signal Setting 1 Online Calibration 2 Signal Correction 3 SS Factor Signal Correction Signal Setting 1 Online Calibration 2 Signal Correction 3 SS Factor Signal Setting 2 Signal Correction 3 SS Factor 4 Wiping Time Wiping Time SS Factor ...

Page 27: ...ulfite solution configured with deionized water and the second point corresponds to air calibration which can be in air or saturated air and water To calibrate after entering the calibration interface you need to wait for a while and after the measured value is stable use to complete the calibration Figure 31 7 2 5 2 Signal Revision The measured dissolved oxygen value can be corrected Figure 32 Si...

Page 28: ...libration of any standard solution enter the conductivity calibration input interface and enter the calibration solution value to be calibrated common standard solutions are 147 0μS cm 1413μS cm and 12 88mS cm the actual can be calibrated according to on site standards Press to enter the conductivity calibration interface and then put the connected electrode into the standard solution After the re...

Page 29: ...measured conductivity value can be corrected Figure 36 7 2 6 3 Electrode constant Set the sensor cell constant Figure 37 Signal Setting 1 Online Calibration 2 Signal Correction 3 Electrode Constant Signal Correction Signal Setting 1 Online Calibration 2 Signal Correction 3 Electrode Constant Electrode Constant ...

Page 30: ...tor for conductivity and total dissolved solids Figure 39 7 2 6 6 Measurement parameters Select the sensor to measure the main parameter Figure 40 Signal Setting 2 Signal Correction 3 Electrode Constant 4 Temp Coefficient Temp Coefficient Signal Setting 3 Electrode Constant 4 Temp Coefficient 5 TDS Coefficient TDS Coefficient Signal Setting 4 Temp Coefficient 5 TDS Coefficient 6 Measure Param Meas...

Page 31: ...emote transmission setting interface Figure 41 7 3 2 Current transmission The engineering value corresponding to 4mA and the engineering value corresponding to 20mA of 4 20 mA output can be set select the number of digits to set the parameter data change data size Jump to 20mA parameter setting save the setting parameters Return to the monitoring interface Return to 4mA parameter setting return to...

Page 32: ...e normally open contact of the high report relay is disconnected and the normally closed contact is closed Select to set parameter data digits Change data size Jump to hysteresis parameter setting save the setting parameters Back to monitoring interface Return to the alarm value parameter setting return to the alarm setting interface Note The unit xx and the data size are automatically switched ac...

Page 33: ... hysteresis When the low alarm relay normally open contact is disconnected the normally closed contact is closed Select to set parameter data digits Change data size Jump to hysteresis parameter setting save the setting parameters Back to monitoring interface Return to the alarm value parameter setting return to the alarm setting interface Note The unit xx and the data size are automatically switc...

Page 34: ...Chapter 7 Configuration Settings www supmea com 27 7 5 Information query interface Version Information Query the current software version Figure 45 Information Inquiry 1 Version Information ...

Page 35: ...re shown in Table 3 Table 3 Communication data and register address Address Data Type Function Code Description Access Rights 0x1100 short 0x03 0x06 Slave address 1 254 W R 0x1101 short 0x03 0x06 Baud rate 0 2400 1 4800 2 9600 default 3 19200 4 57600 W R 0x1102 short 0x03 0x06 Check digit 0 N81 default 1 O81 2 E8 1 N No parity E Even parity O odd parity 8 8 data bits W R 0x2000 short 0x03 High byt...

Page 36: ...R 0x2103 short 0x03 High Alarm Relay Status R 0x2104 short 0x03 0 Disconnect 1 Pull in R Note The above register data transmission adopts big endian mode the high byte is first and the byte order of the floating point type is 4321 Table 4 Sensor Type Correspondence Table Sensor type Data Measurement value 1 2 3 meaning pH 0x0001 Measured value 1 is pH value unit pH Measured value 2 is meaningless ...

Page 37: ...ty value unit ppt Measured value 2 is meaningless value is 0 Measured value 3 is meaningless value is 0 Dissolved oxygen 0x0006 Measured value 1 is the dissolved oxygen value unit mg L Measured value 2 is the dissolved oxygen saturation unit Measured value 3 is meaningless value is 0 Turbidity 0x0007 Measured value 1 is the turbidity value unit NTU Measured value 2 is meaningless value is 0 Measur...

Page 38: ...y device address is 0x01 Device address Function code return byte Return data CRC 0x01 0x03 0x02 0x0001 0x7984 Explanation The current device table address is found to be 0x01 2 Query real time data Device address Function code Start address Number of registers CRC 0x01 0x03 0x2001 0x0009 0xDFCC Data reply device address is 0x01 Device address Function code Return byte Return data CRC 0x01 0x03 0x...

Page 39: ...here are any interference devices such as inverters around and pay attention to stay away from these interference devices or take shielding measures Q3 The controller cannot be calibrated Answer The standard solution is not prepared correctly or the electrode is damaged Question 4 The display area of the screen is that the sensor is not connected A Confirm whether the sensor is disconnected Q5 Wro...

Page 40: ...t charge and all non customized products are guaranteed to be returned and replaced within 7 days Disclaimers During the warranty period product faults caused by the following reasons are not in the scope of Three Guarantees service Product faults caused by improper use by customers Product faults caused by disassembling repairing and refitting the product After sales service commitment We promise...