Chapter 8 Communication Protocol
www.supmea.com
29
0x2001
short
0x03
The type of access sensor
determines the meaning of
measurement values 1, 2,
and 3. Please refer to Table 4
for details.
R
0x2002
float
0x03
Temperature value Unit: °C
R
0x2004
float
0x03
Measured value 1, look up
table 4
R
0x2006
float
0x03
Measured value 2, look up
table 4
R
0x2008
float
0x03
Measured value 3, look up
table 4
R
0x2103
short
0x03
High Alarm Relay Status
R
0x2104
short
0x03
0: Disconnect 1: Pull in
R
Note: The above register data transmission adopts big endian mode, the high
byte is first, and the byte order of the floating point type is 4321
Table 4 Sensor Type Correspondence Table
Sensor type
Data
Measurement value 1, 2, 3 meaning
pH
0x0001
Measured value 1 is pH value, unit: pH
Measured value 2 is meaningless, value is 0
Measured value 3 is meaningless, value is 0
ORP
0x0002
Measured value 1 is the ORP value, unit: mV
Measured value 2 is meaningless, value is 0
Measured value 3 is meaningless, value is 0