Chapter 7 Configuration Settings
www.supmea.com
21
7.2.5.3 Salinity setting:
Sets the salinity value of the current solution. The default value is 0.00ppt.
Figure 33
7.2.5.4 The Pressure setting:
Sets the barometric pressure value at the current altitude. The default is 1standard
atmosphere(760mmHg).
Figure
34
7.2.6 Conductivity Signal Settings
7.2.6.1 Online calibration:
You can perform single-point calibration of any standard solution, enter the
conductivity calibration input interface, and enter the calibration solution value to
be calibrated (common standard solutions are 147.0
μ
S/cm, 1413
μ
S/cm and
12.88mS/cm, the actual can be calibrated according to on-site standards )
,
Press
to
enter the conductivity calibration interface, and then put the connected
electrode into the standard solution. After the reading is stable, press the
--Signal Setting--
1.Online Calibration
2.Signal Correction
3.Salinity Setting
--Salinity Setting--
--Signal Setting--
2.Signal Correction
3.Salinity Setting
4.Atm.Pressure
--Atm.Pressure--