80-0084-00-RevB SunTech Vet30E User Manual | 50
2019-12-04
Appendix A: Service Screens
WARNING:
Do not disassemble the unit. There are no user serviceable parts except for the battery. Refer to qualified
service personnel.
WARNING:
The USB port is only a service port and cannot be used to download data or interface with the monitor.
When updating the software, the monitor cannot be in use and the accessories should not be contacting the patient.
CAUTION:
DO NOT remove unit covers. Doing so may increase the risk of electrical shock. This monitor does not
contain any user serviceable parts. Substitution of a component or accessory different from that supplied may result
in measurement error. Repairs should be undertaken only by personnel trained or authorized by SunTech Medical.
The SunTech Vet30E BP monitor includes a service mode where service information can be accessed, factory default
settings can be restored, and hardware verification checks can be performed. The service screens are accessed by
holding down the power button for approximately 6 seconds when turning the monitor ON. When viewing service
screens, the Start/Stop button LED is white.
No parts on the monitor or its accessories are serviceable.
Service Information
Basic details are logged which includes the number of measurement cycles, user interface firmware version, both
firmware versions for the BP algorithm as well as a count for each status message. Status codes are provided to
assist in troubleshooting during a service call. The image below is provided for example only.
Calibration Check
The pressure transducer in the SunTech Vet30E BP monitor is designed to hold its calibration for many years. Human
NIBP devices require that the maximum static pressure accuracy shall be ± 3mmHg or 2% of the reading, whichever
is greater. This is a stringent requirement and all test equipment must be in excellent working order to properly
perform this test. If you do not have access to this equipment or prefer to have someone else verify the calibration,
the monitor can be sent to SunTech following the procedure described in the Limited Warranty section of this
manual. There may be a charge associated with the verification if the transducer is not out of calibration.