80-0084-00-RevB SunTech Vet30E User Manual | 48
2019-12-04
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
NOTE:
UT is the a.c. mains voltage prior to application of the test level
Conducted RF IEC 61000-
4-6
All device input and
output connections
and cables
3V
10V ISM bands
150kHz to 80MHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
monitor including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter. Minimum separation
distance for higher IMMUNITY TEST
LEVELS shall be calculated using the
following equation.
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m) and E is the
Immunity Test Level in V/m.
𝐸 =
6
𝑑
√𝑃
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a
should be less than the
compliance level in each
frequency range
Rated maximum output
power of transmitter.
Watts (W)
Separation distance according to frequency of transmitter meters (m)
150 kHz to 80 MHz
𝑑 = 1,2√𝑃
80 MHz to 800 MHz
𝑑 = 1,2√𝑃
800 MHz to 2,5 GHz
𝑑 = 2,3√𝑃
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: at 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.