2019-12-04
Replaceable Parts
CAUTION:
Replace user replaceable parts that are broken, worn, missing, damaged, incomplete, or contaminated.
Contact SunTech Medical for service on parts that are not user replaceable and stop using device until repaired.
Failure to repair a damaged product may cause injury to the user and/or patient.
On a routine basis, inspect the monitor, cuffs, sensors, and hoses for cracks, fraying, or kinks. Immediately replace
any damaged part. Device can function without a battery. However, if battery fails to maintain a charge, replace the
battery. Before replacing the battery, make sure the device is turned off and disconnected from patient. Do not
replace the battery without removing power supply from wall. The start/stop button does not remove power from the
system. To replace the battery, use a screwdriver to unscrew the screw and open the battery bay door on the bottom
of the device. Remove battery by disconnecting the electrical connector. Replace battery and reconnect the electrical
connector. Replace battery bay door and secure with screw. Before use, battery must be fully charged.
CAUTION:
Use only approved accessories.
Replace battery only with SunTech part number 17-0026-00.
Expected Service Life
Monitor: 7 years (dependent upon internal BP pump usage)
AccuVet SpO
2
sensor: 3 years
Masimo Patient Cable: 17,520 hours of patient monitoring time
AccuVet temperature probe: 3 years
Lithium-ion Battery: 250 cycles (before capacity falls below 80% of original capacity)
10.4 System Components
Your Vet30E System should contain the following items.
Vet30E System
Standard Qty.
Vet30E System Monitor
1
Patient Hose, 6ft (1.8m)
1