Improper sensor type or application
•
Low perfusion
•
Excessive motion artifact
•
Excessive ambient or strobing light
•
Low battery/not plugged into AC power supply
When troubleshooting a difficulty in obtaining a reading, ensure the following steps are taken:
•
Allow time for reading to stabilize
•
Verify sensor type and check connection to Vet30E
•
Check if blood flow to the sensor site is restricted
•
Check the placement of the sensor and re-apply sensor if needed
•
Replace sensor if defective
•
Shield the sensor from excessive or strobing light
•
Minimize motion at the monitoring site
•
Connect the AC power supply
Refer to the Directions for Use included with the Masimo sensor for additional application information.
7.2 AccuVet Pulse Oximetry
NOTE
: A functional tester cannot be used to assess the accuracy of the pulse oximetry function of the Vet30E.
NOTE
: The recommended site to measure pulse oximetry in horses is the rectum.
Where to Apply the AccuVet Y-lingual SpO
2
Sensor
1.
Clean the sensor before use.
2.
Select the sensor clip that is appropriate for the patient.
3.
Open clip with thumb and a finger. For each side of the clip, push the sensor alignment buttons into the clip
slots until they are fully engaged. Verify that the sensor pads are properly aligned with each other.
4.
Apply the sensor to the patient at the tongue or alternatively at the lip, ear, skin between toes, prepuce or
vulva while the patient is lying down.
Where to Apply the AccuVet Reflectance SpO
2
Sensor
1.
Clean the sensor before use.
2.
Apply disposable, non-rigid covers as needed. Users should ensure sensor covers are used in accordance
with manufacturers’ instructions.
3.
Place sensor in the rectum.