80-0084-00-RevB SunTech Vet30E User Manual | 17
2019-12-04
Clinical Alarms
The SunTech Vet30E allows clinical alarms for all values (SYS, DIA, MAP, HR, Temp, and SpO
2
). Factory setting is
OFF.
•
To change alarm values, select the Settings screen in the Icon Menu, then choose Clinical Alarms.
•
Select the parameter of interest.
•
Touch On, then Set Alarm Values.
•
Touch the numeric range value that you want to adjust. The alarm range cannot be adjusted by dragging the blue
bar. The value on the left side represents the value for which all values below that number will trigger the alarm.
The value on the right side represents the value for which all values above that number will trigger the alarm.
•
Using the keypad, type in the desired value and touch the checkmark.
•
To reset to default clinical alarms, touch Use Defaults. For information on the default clinical alarm values, see
Section 14.
Touch the numbers
on the right or left to
set the alarm limits.
Time
. Set the time.
Date
. Set the date.
Language.
Select desired language.
Temperature.
Select the units for temperature.