CHAPTER SIX
BATTERY CHARGER
ST PS
Sunrise Training and Education Programmes
6:2
2
1
4
Battery Charger
Troubleshooting
3
6
5
RED LIGHT when batteries & mains are connected.
NO LIGHT when batteries & mains are connected.
Lever the fuse carrier out & check the 3A fuse.
Check the mains lead is pushed home into its
sockets. Confirm mains wall socket is OK
Check charging socket for battery voltage. If below
17V change batteries. If zero check reset button,
charging fuse, main fuses & wiring in battery box
(Ch 5).
AMBER LIGHT stays on for longer than 24 hours.
Check the charger plug & socket for moisture or
dirty contacts. If burning is evident change the
socket AND the charger. Perform discharge test on
batteries as they may be worn out or faulty.