Error message list
9. Troubleshooting
Error message
Possible cause
Troubleshooting
Refer to
The splicer failed to
adjust the LED
brightness.
The fiber is loaded in a too
forward position.
Make sure that the right
and left fibers are
placed in the correct
position of the
splicer.
▸
P.2-14
Inserting fibers
into the splicer
Fiber cleave length is too
long.
When cleaving the fiber, make sure
it is loaded in a correct position in the
cleaver and cleave it to a correct
cleave length.
▸
Instruction
manual that
came with the
cleaver you
use.
There is a foreign object in
the screen.
Clean the microscope lens protection
glass and LEDs.
▸
P.3-2
Cleaning LEDs,
P.3-3 Cleaning
lens protection
glass
The fibers are not
placed correctly in
the splicer.
The fiber is not loaded in an
appropriate position.
Make sure that the right
and left fibers are
placed in the correct
position of the splicer.
▸
P.2-14
Inserting fibers
into the splicer
There is a foreign object in
the screen.
Clean the microscope lens protection
glass and LEDs.
▸
P.3-2
Cleaning LEDs,
P.3-3 Cleaning
lens protection
glass
If you encounter any errors during use, please check the error message that appears
and see the error message list below to troubleshoot.
❚
Error message list
LED
Cleave length
Bare fiber
Microscope lens protection glass
NG
NG
NG
OK
OK