Suez Sievers M500 Quick Start Manual Download Page 2

1

DQS 78000-01 MUL Rev. A

© 2020 SUEZ. All Rights Reserved.

Sievers

®

 is a trademark of SUEZ and may be registered in one or more countries.

Sievers M500 

Sievers M500 

Online TOC Analyzer

Online TOC Analyzer

Quick Start Guide

IMPORTANT NOTICE

For detailed instructions, download the 

Sievers M500 Online TOC Analyzer

manual from www.sieversinstruments.com [Customer Support > Sievers 
Analyzers & Instruments > Under “Manuals and Validation Support Packages 
(VSPs),” Sievers M500]..

Installation Instructions

 Step 1 

Unpack and inspect the Analyzer.

 Step 2 

Complete the identi

fi

 cation records.

 Step 3 

Select a location for the Analyzer.

The 

Sievers M500 Online TOC Analyzer

 is designed to be mounted on a wall 

or support stand. It has pre-installed mounts that create a minimum 2.5 cm 
between the back of the Analyzer and the wall for heat dissipation. Allow at 
least 30.5 cm of clearance on the sides, top, and bottom of the Analyzer for 
the plumbing and electrical connections. Additionally, this clearance provides 
for the proper circulation for temperature and humidity control. When selecting 
the location, mount the Analyzer so that the display screen is approximately at 
eye-level.

Select and procure hardware for mounting the Analyzer on a wall or instrument 
rack, based on site-speci

fi

 c circumstances. Mounting hardware must be able to 

support four times the weight of the unit [65.2 kg (144 lbs)]. 

Avoid direct sunlight and extreme temperatures. Operating at elevated 
temperatures greater than 40 °C (104 °F) prevents proper operation, and 
operating at temperatures lower than 5 °C (41 °F), can cause measurement 
errors. 

 Step 4 

Install I/O control wiring and power.

 Step 4 A 

Access wiring areas

(Use ESD protection while working inside the 
Analyzer)

1. 

Unlock and push the release button to unlatch 
and open the door to the Analyzer.

2. 

Remove the electronics cover (Figure 1, at A) 
and I/O Control wiring cover (B) by turning the 
fasteners 1/4 turn to release.

 Step 4 B 

Install I/O wiring.

1. 

Locate the instrument controller board 
(Figure 2, at C) and wire glands (D) inside the 
Analyzer.

2. 

Route the I/O wires into the instrument 
through the wire glands from the outside I/O 
panel, located on the left side of the Analyzer.

3. 

Install the wires into their corresponding terminal blocks, shown below. 
Additionally, see Table 1 and Table 2, at right, for pin information regarding 
alarm relay outputs and 4-20 mA analog current loop outlets. 

Step 5 

Install power.

This step should only be performed by a quali

fi

 ed 

electrician.

To install the AC conduit:

1. 

Pull the AC line cable through the AC mains 
wire gland on the I/O panel of the enclosure 
(Figure 3).

2. 

Strip back the outer insulation by ~3 1/2 
inches.

Warnings

See the separate “Warnings” document for additional important information.

The iOS system and vial ports contain sharp needles 

designed to pierce the septa of sample 

vials. Do not put 

fi

 ngers or inappropriate materials into the iOS system or vial port.

To avoid false TOC readings and possible damage 

to the Analyzer, always make sure 

the sample inlet is open and the DI water reservoir is 

fi

 lled BEFORE turning on the Analyzer. 

Always close the door of the iOS System when not sampling from a vial.

RISK GROUP 3 Warning:

 UV emitted from this product. Avoid eye and skin exposure to 

unshielded product.

English

English

This symbol on the instrument indicates that the product does contain 
restricted substances included in China RoHS II. Please refer to the 
following table.

Water Technologies & Solutions

     Figure 3 - Power cable path 

Figure 1 - Covers 

Figure 2 - I/O Wiring Areas 

Table 1:
Alarm Relay 
Outputs (J1)

Pin Number

Signal

Function

1

ALARM_1_NC

Relay 1 Normally Closed contact

2

ALARM_1_COM

Relay 1 Common contact

3

ALARM_1_NO

Relay 1 Normally Open contact

4

ALARM_2_NC

Relay 2 Normally Closed contact

5

ALARM_2_COM

Relay 2 Normally Common contact

6

ALARM_2_NO

Relay 2 Normally Open contact

7

ALARM_3_NC

Relay 3 Normally Closed contact

8

ALARM_3_COM

Relay 3 Normally Common contact

9

ALARM_3_NO

Relay 3 Normally Open contact

10

ALARM_4_NC

Relay 4 Normally Closed contact

11

ALARM_4_COM

Relay 4 Normally Common contact

12

ALARM_4_NO

Relay 4 Normally Open contact

Table 2:
4-20 mA 
Analog Current 
Loop Outputs 
(J4)

Pin Number

Output

1

#1   4-20 mA  (Return) -

2

#1   4-20 mA  (Out) +

3

#2   4-20 mA  (Return) -

4

#2   4-20 mA  (Out) +

5

#3   4-20 mA  (Return) -

6

#3   4-20 mA  (Out) +

Please Note: This product contains a UV Lamp (with Hg) which meets the exemption de

fi

 ned 

in Annex III 4(a) of RoHS Directive 2011/65/EC.

g

Summary of Contents for Sievers M500

Page 1: ...DQS 78000 01 MUL Sievers M500 Quick Start Guide Total Organic Carbon TOC Analyzer Sievers M500...

Page 2: ...the outside I O panel located on the left side of the Analyzer 3 Install the wires into their corresponding terminal blocks shown below Additionally see Table 1 and Table 2 at right for pin informati...

Page 3: ...d crimp a ring terminal lug for M4 or 8 stud onto the earth lead ground conductor 2 Locate the terminal block on the back of the rotary AC power switch and connect the AC main wires to the front Figur...

Page 4: ...al izquierdo del analizador 3 Instalar los cables en los bloques de terminales correspondientes que se muestran m s abajo Asimismo v anse los cuadros 1 y 2 a la derecha para obtener informaci n de los...

Page 5: ...a parte posterior del interruptor de corriente CA giratorio y conectar los cables del suministro de CA a la parte frontal figura 4 3 Conectar el cable de fase en general marr n o negro al terminal n 5...

Page 6: ...ievers M500 TOC 2 5 cm 30 5 cm 4 65 2 kg 144 40 C 104 F 5 C 41 F 4 4A ESD 1 2 1 4 1 A 1 B 4B 1 2 C 2 D 2 3 4 20 mA 1 2 5 1 3 2 3 5 9 cm L26 iOS 72 iOS RoHS II Water Technologies Solutions Water Techno...

Page 7: ...be registered in one or more countries SSE 72 f f SSE f f SSE SSE 2 SSE 2 C F SVLJ L L 9 C F P 56 86 P 7 8 6 1R FP FP FP NJ 3 DF D E J Q 0RGEXV 7 3 3 3URILQHW 3URILEXV 6 FP 6 FP 56 1 M4 8 2 4 3 5 A 4...

Page 8: ...2 3 Sievers M500 TOC 2 5cm 30 5 cm 4 40 5 4 I O 4A 1 2 1 4 1 A I O B 4B I O 1 2 C D 2 I O I O 3 4 20 mA 1 2 5 AC 1 I O AC AC 3 2 3 1 2 9 cm 2 TOC DI iOS 3 UV China RoHS II Water Technologies Solution...

Page 9: ...ountries t 72 SSE s s SSE s s SSE 2 SSE 2 SSE P PLQ P PLQ F SVLJ 3 L L 9 F P 56 86 P 7 8 6 1R FP FP FP NJ 3 DF D E J Q 0RGEXV 7 3 3 3URILQHW 3URILEXV t 6 FP 6 FP 1 2 AC AC 4 3 No 5 A No 4 B 4 AC 5 6 M...

Page 10: ...QqB B P UPB B B B B B B PB NB B B BSB BT B B B B BVOTRB cB B B B B B B B P WB B B B B B B B B B P ceB B SPB BkQqB B BceB B B B BceB B B J BUKP TPB B B UBSQTBJ B KB B P B B B B B B P quB B B B B B B B...

Page 11: ...aG U G GOYYUW PSGkaGY_U G OXXUZ P G X UZG OZ G P pwG _WYUXX V V V V G G GYU V Gno t G jwVpwG Gw O PSGw O P G G G GXW U G G G G G G G G G U WUWXGTG_WWG zV WUWW G zV GXLG G G GXUWGLGyzk SPB B B B O B B...

Page 12: ...r situ sur le c t gauche de l analyseur 3 Installez les c bles dans les r partiteurs correspondants comme indiqu ci dessous En outre voir le tableau 1 et le tableau 2 droite pour les informations sur...

Page 13: ...se de bague terminale pour le goujon M4 ou 8 sur le conducteur de terre 2 Localisez le r partiteur l arri re de l interrupteur rotatif d alimentation AC et connectez les c bles principaux du AC l avan...

Page 14: ...der u eren E A Tafel die sich auf der linken Seite des Analysators befindet durch die Kabelverschraubungen in das Ger t 3 Installieren Sie die Kabel in die entsprechenden Klemmenbl cke wie unten darge...

Page 15: ...R ckseite des Netzschalters und schlie en Sie die Stromversorgungsanschl sse an der Vorderseite an Abbildung 4 3 Schlie en Sie die Phase typischerweise braun oder schwarz an Klemme 5 A und den Neutral...

Page 16: ...esterno situato sul lato sinistro dell Analizzatore 3 Installare i cavi nelle morsettiere corrispondenti mostrate di seguito Inoltre vedere Tabella 1 e Tabella 2 a destra per informazioni sui pin rigu...

Page 17: ...uare la morsettiera sul retro della manopola CA e collegare i cavi CA nella parte anteriore Figura 4 3 Collegare il filo di rete in genere di colore marrone o nero al morsetto n 5 A e il filo neutro i...

Page 18: ...fio do painel externo de E S localizado esquerda do analisador 3 Instale os fios nos terminais correspondentes mostrados a seguir Consulte tamb m a Tabela 1 e a Tabela 2 direita para obter as informa...

Page 19: ...o 2 Localize o terminal na parte de tr s do interruptor rotativo de CA e conecte os fios principais de CA parte da frente Figura 4 3 Conecte o fio da linha de alimenta o normalmente marrom ou preto ao...

Page 20: ...inal This symbol indicates the protective earth terminal ground for the Analyzer ground for the Analyzer Risk Group 3 Warning UV emitted from this product Avoid eye Risk Group 3 Warning UV emitted fro...

Page 21: ...running Make sure the DI water reservoir is full particularly when running samples with high TOC or high salt concentrations After running samples with high TOC or high salt concentrations After runni...

Page 22: ...ocar nada dentro del analizador mientras se observa la bomba de muestras dentro del analizador mientras se observa la bomba de muestras durante una inspecci n visual durante una inspecci n visual Espa...

Page 23: ...ved Sievers is a trademark of SUEZ and may be registered in one or more countries TOC TOC TOC TOC TOC TOC WEEE 2012 19 EU WEEE 2012 19 EU 2006 66 EC 2006 66 EC TOC TOC iOS iOS RoHS II RoHS 2011 65 EC...

Page 24: ...All Rights Reserved Sievers is a trademark of SUEZ and may be registered in one or more countries iOS iOS iOS iOS SUEZ SUEZ WEEE WEEE 2012 19 EU 2012 19 EU 6 1 6 1 iOS iOS iOS iOS 30 30 iOS iOS 3 3 Po...

Page 25: ...tries 2 2 SUEZ SUEZ WEEE EU WEEE EU 2012 19 EU 2012 19 EU UV UV UV UV UV UV UV UV 15 cm 15 cm 1 1 iOS iOS iOS iOS 30 30 iOS iOS 3 3 A A MSDS MSDS I I AC AC UV UV Sievers M500 TOC Sievers M500 TOC pH 3...

Page 26: ...B PB B B B B B B B PB B B B B B B B B B B B B B B B B B B B B P B P B B B BceB B B B B B B BceB B B B B B B P B B B P B B B B B B B B B B B B B B B B B B B B B B B B P B B B B B B P B B B B B B B B B...

Page 27: ...tout choc potentiellement dangereux ne touchez Pour viter tout choc potentiellement dangereux ne touchez rien l int rieur de l analyseur lorsque vous observez la pompe rien l int rieur de l analyseur...

Page 28: ...zu gew hrleisten muss die Schutzerde der Stromversorgung sicher mit dem Geh use die Schutzerde der Stromversorgung sicher mit dem Geh use verbunden sein verbunden sein Um einen m glicherweise gef hrli...

Page 29: ...lloggiamento Per evitare scosse potenzialmente pericolose non toccare Per evitare scosse potenzialmente pericolose non toccare l interno dell Analizzatore durante il controllo visivo della pompa di l...

Page 30: ...que em nada dentro do analisador enquanto examina a bomba de amostras dentro do analisador enquanto examina a bomba de amostras Portugu s Avisos A l mpada UV cont m merc rio e pode ser considerada mat...

Page 31: ......

Page 32: ...2020 SUEZ All Rights Reserved Sievers is a trademark of SUEZ and may be registered in one or more countries DQS 78000 01 MUL Rev A 6060 Spine Road Boulder CO 80301 3687 USA www sieversinstruments com...

Reviews: