background image

Sturtevant Richmont

 -  P/N 857285 - Issue Date 100407 - Page 10 of 14

Tool Display Note

• 

At this point the tool LCD should be displaying “0” in the lower right corner of the LCD display.

• 

The tool display will use both the main display and the lower right-hand segment to show key informa-
tion needed during these processes.

Procedure - Calibration Adjustment

1.  Install drive square on torque tester. Place wrench onto torque tester. Apply three 100% CW FS loads 

to ExS 3xxx wrench. See Pages 14 - 16 of this document for correct loads.

2.  Remove the torque load from wrench.  Remove wrench from tester, and hold in vertical position. Press 

DOWN arrow key to save the zero setting. Replace wrench on tester.

3.  Press the Up arrow key until 100 appears in the lower right corner of the LCD display. Apply 100% CW 

FS load to the unit.  When exactly 100% CW FS torque is applied, press the DOWN arrow key. Unload 
the tool.

4.  Press the Up arrow key until 90 appears in the lower right corner of the LCD display. Apply 90% CW FS 

load to the unit.  When exactly 90% CW FS torque is applied, press the DOWN arrow key. Unload the 
tool.

5.  Press the Up arrow key until 80 appears in the lower right corner of the LCD display.  Apply 80% CW 

FS load to the unit.   When exactly 80% CW FS torque is applied, press the DOWN arrow key. Unload 
the tool.

6.  Press the Up arrow key until 70 appears in the lower right corner of the LCD display. Apply 70% CW FS 

load to the unit.  When exactly 70% CW FS torque is applied, press the DOWN arrow key. Unload the 
tool.

7.  Press the Up arrow key until 60 appears in the lower right corner of the LCD display. Apply 60% CW FS 

load to the unit. When exactly 60% CW FS torque is applied, press the DOWN arrow key. Unload the 
tool.

8.  Press the Up arrow key until 50 appears in the lower right corner of the LCD display. Apply 50% CW FS 

load to the unit.  When exactly 50% CW FS torque is applied, press the DOWN arrow key. Unload the 
tool.

9.  Press the Up arrow key until 40 appears in the lower right corner of the LCD display. Apply 40% CW FS 

load to the unit. When exactly 40% CW FS torque is applied, press the DOWN arrow key. Unload the 
tool.

10. Press the Up arrow key until 30 appears in the lower right corner of the LCD display. Apply 30% CW FS 

load to the unit.  When exactly 30% CW FS torque is applied, press the DOWN arrow key. Unload the 
tool.

11. Press the Up arrow key until 20 appears in the lower right corner of the LCD display.  Apply 20% CW 

FS load to the unit. When exactly 20% CW FS torque is applied, press the DOWN arrow key. Unload 
the tool.

12. Press the Up arrow key until 10 appears in the lower right corner of the LCD display.  Apply 10% CW 

FS load to the unit. When exactly 10% CW FS torque is applied, press the DOWN arrow key. Unload 
the tool.

Summary of Contents for exacta 2 3000 series

Page 1: ...nd Battery Chargers These tools were designed to use only and solely NiMH batteries having specific dimensions and operating characteristics Within NAFTA countries the only batteries that are to be us...

Page 2: ...At that time the display will show the message bAtt and the ability to operate the wrench will be lost Readings stored prior to that point will remain in memory The tool will remain inoperable until...

Page 3: ...tion here due to differences in the fonts used Target When the triangular icon beneath this is illuminated it means that the target value for the torque specification is displayed in the main section...

Page 4: ...specific Task Numbered from 00 to 50 Record 00 is designed to give the user the ability to use the specifications for the active task on a fastener without recording the result UOM Currently active Un...

Page 5: ...ted Up Navigate among tasks and records Increments count of displayed task or record by one Down Navigate among tasks and records Decrements count of displayed task or record by one Display During Ope...

Page 6: ...the Exacta Plus software is being used to communicate with it avoid touching the buttons on the tool and do not disable the power Allow the software to perform the communications and complete them bef...

Page 7: ...records for that task have been obtained Once the last fastener in a task has been completed the information will be displayed normally then the screen will display SEt in the main section Use the Up...

Page 8: ...but the use of a mechanical loader is still recommended The load applied during testing must be at the loading point on the grip indicated by the plain ring at the middle of the grip It is imperative...

Page 9: ...diagnostic D Command into the terminal and press enter The Terminal display will dis play DiaOn You are now in diagnostic mode 8 Type the M enter The capacity of the tool should be seen on the displa...

Page 10: ...ress the DOWN arrow key Unload the tool 6 Press the Up arrow key until 70 appears in the lower right corner of the LCD display Apply 70 CW FS load to the unit When exactly 70 CW FS torque is applied p...

Page 11: ...load to the unit When exactly 60 CCW FS torque is applied press the DOWN arrow key Unload the tool 20 Press the Up arrow key until 50 appears in the lower right corner of the LCD display Apply 50 CCW...

Page 12: ...300 303 297 75 foot pounds Capacity in Ft Lb FS Torque Value in In Lb Tol Tol 75 10 90 90 9 89 1 20 180 181 8 178 2 30 270 272 7 267 3 40 360 363 6 356 4 50 450 454 5 445 5 60 540 545 4 534 6 70 630...

Page 13: ...4 30 1440 1454 4 1425 6 40 1920 1939 2 1900 8 50 2400 2424 2376 60 2880 2908 8 2851 2 70 3360 3393 6 3326 4 80 3840 3878 4 3801 6 90 4320 4363 2 4276 8 100 4800 4848 4752 600 foot pounds Capacity in F...

Page 14: ...to the sensor including damage of a level that necessitates replacement We strongly recommend that the tool never be used beyond its rated capacity and that if the tool is inadvertently overloaded it...