11. Specifications
Model
SA-300VMA
SA-300VMA-R
Chamber Capacity (L)
50
Maximum Instrument Length
(mm)
600
Maximum Load (unpouched) (g)
10,000
Maximum Load (pouched) (g)
1,000
External Dimensions (mm)
980(D) x 600(W) x 600 (H)
Chamber Size (mm)
300 Diameter x 710 Depth
Gross Weight (kg)
104
Voltage/Wattage (Heater)
220V-240V AC, 50/60Hz, 2900W
Fuses
20A x 2, No Fuse (circuit) Breaker
Water Capacity per Cycle (ml)
1000 ~1200
Sterilization Temperature (°C)
121 / 134
Working Environment
Indoor use;
Under 1,000m (altitude);
Temperature 5°C to 40°C;
Relative Humidity 80%RH@31°C to Relative Humidity
50%RH@40°C;
Voltage fluctuation ±10 %;
Transient overvoltages category II;
Pollution degree 2
Transportation Conditions
-10°C to 70°C, 10%RH to 90%RH
Storage Conditions
-10°C to 50°C, 10%RH - 70%RH
Over Pressure Protection
Safety valve 2.55 bar(2.6 kgf/cm²)
Over Pressure Indication
Pressure protection switch with warning LED
Over Temperature Indication
Independent temperature protection device with warning
LED
Water Level Indication
Water level sensor in reservoir and chamber with warning
LED
Door Lock Indication
Micro switch sensor with warning LED
Pressure Display
Analog pressure gauge
Function Display
LED
Recorder
─
Yes
Program Selections
Wrapped/unwrapped, Dry/no dry, Liquid.
Temperature Selection
121°C/134°C
Drying Function
0, 15, 30, 35, 40 min. (Re-Dry: 10 min.)
Additional controls
PRION, B.D. Test, Pump Test, Liquid, Reset/Stop, Vacuum
release, Emergency
Standard Accessories
Basket x 2; Heater cover; 2000cc bottle x 1 pcs
Life Time