7.6 Message and troubleshooting (for recorder)
Symptom
Solution
No display
- Turn on the main switch.
- Plug the power cable.
- Check if bad connection.
- Contact local distributor for service
E101: Temp. Sensor
fail
- Contact local distributor for service
E103: Pressure sensor
fail
- Contact local distributor for service
E105: No USB memory
- Insert a USB flash drive.
- Detach and re-insert memory card again
E107: USB is full
The data stored on the medium have reached its
maximum limit, please backup the storage medium onto a
secure zone immediately, and then followed the procedure
“7.5.2.4 Download” to download the last record.
E108: USB formatter
error
- Re-format the USB flash drive.
- Replace another USB flash drive.
- Contact local distributor for service
E109: EEPRM fail
- Replace another USB flash drive.
- Contact local distributor for service
E125: Keyboard fail
Contact local distributor for service
E135: Abnormal of the
cycle
- Check the power supply to the Autoclave.
- Contact local distributor for service
E155: Password fail
Authorized person only. This function is reserved for
diagnostic purpose by the manufacturer, users are
password protected. Contact local distributor or
manufacturer for information.
Caution:
Contact local distributor for service if encountered any other problems. Do not
attempt to disassemble the sterilizer or recorder by yourself. Failure to do so
could result in serious injury or damage to the unit.