29
7. Recorder (SA-300VMA-R)
7.1 Intended use
This paperless Recorder records the sterilization temperature, theoretical steam pressure and
real time information during each cycle as an optional method of printer. It records the specified
information onto an USB flash drive, in the Excel csv (Comma Separated Values) format which
compatible with Microsoft® Office Excel.
Note:
Insert a formatted USB flash drive before commencing a sterilization cycle.
Note:
Microsoft Office Excel are registered trademarks of Microsoft Inc. Microsoft is a
registered trademark.
Note:
A minimum 20 minutes is recommended between each sterilization cycles, and be
sure of “Data transfer complete” has been completed to a USB before starting a new
sterilization cycle.
7.2 Recording contents
Description of recording contents:
Two type of recording files will be created for each sterilization cycle. One is Detail
Contents in the format of “YYMMDD_HHMMSS.csv”, and the other is the Summary
Contents in the format of “
S
YYMMDD_HHMMSS.csv”.