www.stryker.com
6500-509-001 REV B
17
Summary of Safety Precautions
CAUTION (CONTINUED)
• The weight of the equipment in the base storage net (if equipped) must not exceed 20 pounds (9 kg).
• Be careful when retracting the base to avoid damaging items stored in the base storage net.
• The weight of the equipment in the head end storage flat (if equipped) must not exceed 40 pounds (18 kg).
• DO NOT STEAM CLEAN OR ULTRASONICALLY CLEAN THE UNIT.
• Maximum water temperature should not exceed 180°F/82°C.
• Maximum air dry temperature (cart washers) should not exceed 240°F/115°C.
• Maximum water pressure should not exceed 1500 psi/130.5 bar. If a hand held wand is being used to wash the
unit, the pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.
• Towel dry all casters and interface points.
• Failure to comply with these instructions may invalidate any/all warranties.
• Always remove the battery before washing the cot.
• A preventative maintenance program should be established for all Stryker EMS equipment. Preventative
maintenance may need to be performed more frequently based on the usage level of the product. Close attention
should be given to safety features including, but not limited to:
• Hydraulic power mechanism
• All electrical controls return to off or neutral position when released.
For additional maintenance information, see the preventative maintenance information.
• Improper maintenance can cause injury or damage to the product. Maintain the cot as described in this manual.
Use only Stryker approved parts and maintenance procedures. Using unapproved parts and procedures could
cause unpredictable operation and/or injury and will void the product warranty (see
).
• Failure to use authorized parts, lubricants, etc. could cause damage to the cot and will void the warranty of the
product.
• Hydraulic lines, hoses, and connections can fail or loosen due to physical damage, kinks, age, and exposure.
Check hoses and lines regularly to avoid damage to the cot. Check and tighten loose connections.
• Do not tip the cot onto its load wheels and actuate the product as this will allow air to enter the hydraulic system.
• Do not lubricate the bearings in the X-frame as it will degrade the performance of the cot and may void its warranty
• The cot retaining post comes preconfigured for an X-frame cot, if the fastener has been configured for an H-frame
style cot, the cot retaining post must be adjusted to accommodate the fastener.
Summary of Contents for Power-PRO XT 6500
Page 2: ......
Page 90: ...90 6500 509 001 REV B www stryker com Return To Table of Contents Cot Assembly ...
Page 91: ...www stryker com 6500 509 001 REV B 91 Return To Table of Contents Cot Assembly ...
Page 92: ...92 6500 509 001 REV B www stryker com Return To Table of Contents Cot Assembly ...
Page 93: ...www stryker com 6500 509 001 REV B 93 Return To Table of Contents Cot Assembly ...
Page 97: ...www stryker com 6500 509 001 REV B 97 Return To Table of Contents Base Assembly ...
Page 98: ...98 6500 509 001 REV B www stryker com Return To Table of Contents Base Assembly ...
Page 106: ...106 6500 509 001 REV B www stryker com Return To Table of Contents Notes ...
Page 118: ...118 6500 509 001 REV B www stryker com Return To Table of Contents Foot End Assembly ...
Page 124: ...124 6500 509 001 REV B www stryker com Return To Table of Contents Optional Gatch Assembly ...
Page 171: ......