Preventive maintenance
At a minimum, make sure that all items listed during annual preventive maintenance for all Stryker Medical products.
Preventive maintenance is to be completed by a qualified service technician.
Inspect all of the following items:
Power cord and plug for fraying
Condition of covers and push handle for damage
Hose ports are operational
Ground chain attached
LCD is not cracked
Wheels for smooth operation
Rear caster wheels for free swivel action
Both rear wheels lock securely when the brake is applied
Front and rear wheels are not loose or wobbly
Battery backup functional
Alarm system - visual and audible
LCD functional
Touch screen functional
Water temperature and flow verification
Probe resistance
Clear RFU codes
Ground impedance not more than 100mΩ (millohms)
Current leakage not more than 300 (microamps)
Verify the integrity of all clamps and clamped joints located in the air elimination circuit.
Replace the following on an annual basis:
Replace the 9 volt battery
Replace the condenser inlet filter
Replace the Air eliminator hose
Product Serial Number:
Completed by:
Date: