www.stryker.com
8001-009-001 REV G
43
Disinfecting
Disinfecting the internal water circuit and hoses (Continued)
Note:
Do not rinse the reservoir.
20. Pull up on the controller drain plug (
) to open the drain.
A
Figure 31: Drain plug
21. Make sure that all 3 hoses remain raised above the connection ports for draining.
22. Allow the controller and hoses to drain for a minimum of two minutes.
23. Push down on the controller drain plug to close the drain.
24. When the controller and hoses are drained, continue to
Rinsing the internal water circuit and hoses on page 43
.
Rinsing the internal water circuit and hoses
1.
Using appropriate measuring equipment, fill the empty reservoir with 1 gallons (3.8 L) of sterile-distilled water.
2.
Place the reservoir into the controller.
3.
Connect the bottom hose end to the hydraulic connector in the lid of the reservoir.
Figure 32: Bottom hose end in the lid of the reservoir
4.
Plug the power cord into a wall outlet.
5.
Press and hold the
Stand-by button.
6.
Tap the Manual mode icon.