Introduction
Product alarms (Continued)
Icon, yellow
Name
Alarm priority
and delay
Message
Therapy
interrupted
Check
Check patient
probe (A or B)
Medium
Abnormal change
in patient
temperature
Yes
Check probe
condition,
location, and
connections
Probe or adapter
malfunction (A or
B)
Medium, 30
second delay
No temperature
signal detected.
Yes
Check probe or
adapter cable
condition,
location, and
connections
Adapter cable
disconnected (A
or B)
Medium, 30
second delay
Adapter cable not
detected
Yes
Reinsert the
adapter cable. If
damaged,
replace the
adapter cable
Patient
temperature
deviation
Medium
Patient
temperature is ±
0.5° C (0.9° F)
outside of target
temperature
(Only will appear
after the initial
target is reached.)
No
Check patient
condition,
placement of
thermal transfer
devices, and all
connections
Normothermia
deviation
Low
Patient
temperature is
outside of 36.0° C
(96.8° F) to 38.0°
C (100.4° F)
No
Check patient
condition,
placement of
thermal transfer
devices, and all
connections
Therapy pause
Medium
Therapy is
currently paused
Yes
To resume, press
and hold the play
/ pause for 2
seconds
Battery low
Low
Battery is low
No
Maintenance is
recommended. If
battery is not
replaced, the
product may not
function on the
next startup.
Patient
temperature
output (A or B)
Low
Patient
temperature
output is
inaccurate on the
external device,
or outside
supported range
No
Check output
adapter cable
connection. Tap
Confirm to
reactivate the
output port.