www.stryker.com
8001-009-001 REV G
27
Operation
Filling the reservoir with sterile distilled water (Continued)
CAUTION
•
Always use sterile distilled water or water that has been passed through a filter less than or equal to 0.22 microns
with this product.
•
Always fill the reservoir with room temperature sterile distilled water to reduce the risk of burn.
•
Do not overfill the reservoir to avoid the risk of water spillage and fall.
To fill the removable reservoir with sterile distilled water:
1.
See
Removing and replacing the reservoir on page 26
2.
Fill the reservoir with five liters of sterile distilled water. Do not fill past the top fill line to avoid water overflow (
).
A
Figure 16: Reservoir fill lines
Selecting and setting the primary probe
A patient probe displays when present, stable, and confirmed. The choice of Probe A or Probe B is highlighted when you
insert the cable into port A and port B. If you only insert one cable, the active port displays.
1.
Tap the Settings button.
a.
In the standby mode, tap the Back button to display the edit settings screen.
b.
In an active therapy mode, tap the Next button.
2.
Tap Select Probe to display the Select Primary Probe (Probe A or Probe B) screen. If both probes are present, the
default is probe A.
3.
Tap A or B, if applicable.
4.
Tap Confirm.
Notes
•
The message “Probe stabilization in progress... Please wait” is displayed.
•
When you initially select a probe (A or B), detected is checked. When stabilized, the Ready check is displayed.
•
If the probe is not stabilized within three minutes, the message “Probe stabilization error” will appear. Tap the
Help button for more detail.
•
You can select help at any time to display help with the current screen or icon descriptions.