Introduction
Specifications (Continued)
Patient temperature
Control setting range
89.6° - 100.4° F
32.0° - 38.0° C
Control accuracy
±0.1° C (32° - 38° C)
Measurement accuracy
±0.3° C (25.0° - 45.0° C)
±0.4° C (0.0° C - 24.9° C, 45.1° C - 50.0° C)
Display / resolution setting
0.1° C
Display range
32.0° - 122.0° F
0.0° - 50.0° C
Default setting
98.6° F
37.0° C
Controller
Heater capacity, maximum
500 watts
Circulating fluid
Sterile distilled water or water that has been passed through a filter less than or
equal to 0.22 microns with this product
Battery
9V Lithium
Alarm tone range
75 - 85 dBA per standard IEC 60601-1-8
Water flow rate in each hose port
Typical 1.2 lpm
Refrigerant type
R134a
Power cord length
14 to 15 feet
4.2 - 4.5 meters
Clinical thermometer
Direct mode
Equipment Class
Class I
Rated for continuous operation
Note:
The controller takes approximately 9 minutes to heat from 23.0±2° C (73.4° F) to 37.0° C (98.6° F) when not
connected to a patient. Time will vary when connected to a patient.
Stryker reserves the right to change specifications without notice.
For more information about thermal transfer devices, cables, or probes, see the manufacturer’s instructions for use.
Environmental
conditions
Operation
Storage
Transporta tion
Ambient temperature
89.9°F
(32.2°C)
59°F
(15°C)
140 °F
(60 °C)
-40°F
(-40°C)
140 °F
(60 °C)
-20°F
(-29°C)
Relative humidity (non-
condensing)
75 %
30 %
95 %
10 %
95 %
25 %
Atmospheric pressure
1060 hPa
700 hPa
1060 hPa
500 hPa
Not applicable