Introduction
Product alarms (Continued)
Icon, yellow
Name
Alarm priority
and delay
Message
Therapy
interrupted
Check
Water
temperature
deviation
Medium
Water
temperature is
±0.8° C (1.4° F)
outside of target
temperature
No
Temporary
condition upon
startup, addition
of thermal
transfer device,
or addition of
water
No water
Medium, 20
second delay
No water
Yes
Check for leaks
Add a minimum
of 2 liters of
water
No water flow
Medium, 20
second delay
No flow detected
Yes
Check for leaks
and obstructions
at connections,
hoses, and
thermal transfer
devices
Check water flow
on any port
Medium, 60
second delay
Reduced flow
detected
No
Tap Confirm, if
the water port
was removed
intentionally
Check for leaks
and obstructions
at connections,
hoses, and
thermal transfer
devices