Disinfecting
Disinfecting the product
WARNING
•
Do not clean, service, or perform maintenance while the product is in use.
•
Do not steam clean, hose off, or ultrasonically clean the product. Use of these methods of cleaning is not
recommended and may void the product’s warranty.
Recommended disinfectants
•
Quaternaries (active ingredient - ammonium chloride)
•
Phenolic disinfectant (active ingredient - o-phenylphenol)
•
Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water which equals 520 ppm available
chlorine (40 mL of a 5.25% bleach solution per 4000 mL water))
•
70% isopropyl alcohol
Recommended disinfection method
1.
Follow the disinfectant solution manufacturer’s dilution recommendations.
2.
Hand wash all surfaces of the product with a disinfectant solution.
3.
Avoid over-saturation and make sure that the product does not stay wet longer than the chemical manufacturer’s
guidelines for proper disinfecting.
4.
Dry thoroughly. Do not replace the mattress on the product until the product is dry.
5.
Disinfect the
Velcro®
after every use. Saturate the
Velcro®
with disinfectant, rinse with water, and allow the
disinfectant to evaporate (appropriate disinfectant is determined by the facility).
6.
Check functionality before you return the product to service.
•
Raise and lower the product
•
Lock and unlock the brake/steer pedal in both positions
•
Latch and unlatch the siderails
•
Raise and lower the Fowler backrest
•
Raise and lower the gatch
•
Make sure all components have proper lubrication
•
Make sure all labels are intact
Note:
Some cleaning agents are corrosive in nature and may cause damage to the product if you use them
improperly. If you do not rinse and dry the product well, a corrosive residue may be left on the surface of the
product that could cause premature corrosion of critical components. Failure to follow these cleaning instructions
may void your warranty.
Disinfecting the mattress
These instructions provide recommended disinfecting methods for the mattress.
WARNING
Always disinfect the mattress between patients. Failure to do so could result in cross-contamination and infection.
Recommended disinfectants:
•
Quaternaries (active ingredient - ammonium chloride)
•
Phenolic disinfectant (active ingredient - o-phenylphenol)
•
Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water which equals 520 ppm available
chlorine (40 mL of a 5.25% bleach solution per 4000 mL water))
Summary of Contents for 747
Page 2: ...sample text ...
Page 3: ...2017 05 A 2 0747 109 005 REV A www stryker com Transport Stretcher 747 748 Operations Manual ...
Page 4: ...sample text ...
Page 6: ...sample text ...
Page 38: ...sample text ...
Page 40: ...sample text ...
Page 72: ...sample text ...
Page 74: ...sample text ...
Page 76: ...sample text ...
Page 109: ...2017 05 A 2 0747 109 005 REV A www stryker com Camilla de transporte 747 748 Manual de uso ...
Page 110: ...sample text ...
Page 112: ...sample text ...
Page 144: ...sample text ...
Page 146: ...sample text ...
Page 148: ...sample text ...
Page 180: ...sample text ...
Page 181: ...2017 05 A 2 0747 109 005 REV A www stryker com Barella di trasporto 747 748 Manuale d uso ...
Page 182: ...sample text ...
Page 184: ...sample text ...
Page 216: ...sample text ...
Page 218: ...sample text ...
Page 220: ...sample text ...
Page 252: ...sample text ...
Page 253: ...2017 05 A 2 0747 109 005 REV A www stryker com Transportbåre 747 748 Betjeningsmanual ...
Page 254: ...sample text ...
Page 256: ...sample text ...
Page 288: ...sample text ...
Page 289: ...2017 05 A 2 0747 109 005 REV A www stryker com Transportsjuksäng 747 748 Användarhandbok ...
Page 290: ...sample text ...
Page 292: ...sample text ...
Page 324: ...sample text ...
Page 326: ...sample text ...
Page 328: ...sample text ...
Page 361: ...2017 05 A 2 0747 109 005 REV A www stryker com 수송 운반차 747 748 작동 설명서 ...
Page 362: ...sample text ...
Page 364: ...sample text ...
Page 370: ...한국어 KO 11 6 0747 109 005 REV A www stryker com 안전 예방 조치 요약 압착 지점 그림 11 1 압착 지점 ...
Page 395: ...2017 05 A 2 0747 109 005 REV A www stryker com 转运担架床 747 748 操作手册 ...
Page 396: ...sample text ...
Page 398: ...sample text ...