
page 1–4
Safety Tips and Guidelines
Before operating the 2030, it is important to read and understand all information in this manual. Carefully read
and strictly follow the safety guidelines listed on this page.
It is important that all users have been trained and educated on the inherent hazards associated with the
usage of electric beds.
WARNING
The 2030 is not intended for use with patients less than two years of age.
Powered bed mechanisms can cause serious injury. Operate bed only when all persons are clear of the
mechanisms.
To help reduce the number and severity of falls by patients, always leave the bed in the lowest position
when the patient is unattended.
Leave the siderails fully up and locked when the patient is unattended. When raising the siderails, listen
for the ”click” that indicates the siderail has locked in the up position. Pull firmly on the siderail to ensure
it is locked into position.
Siderails are not intended to be a patient restraint device. It is the responsibility of the attending medical
personnel to determine the degree of restraint necessary to ensure a patient will remain safely in bed.
Always keep the caster brakes applied when a patient is on the bed (except during transport). Serious
injury could result if the bed moves while a patient is getting in or out of bed. After the brake pedal is
applied, push on the bed to ensure the brakes are locked. When moving the bed, toggle the steer pedal
to put the bed in the steer mode. This locks the swivel motion of the right foot end caster and makes the
bed easier to move.
When large spills occur in the area of the circuit boards, 110 volt cables and motors, immediately unplug
the bed power cord from the wall socket. Remove the patient from the bed and clean up the fluid. Have
maintenance completely check the bed. Fluids can affect the operational capabilities of any electrical
product. DO NOT put the bed back into service until it is completely dry and has been thoroughly tested
for safe operation.
Do not steam clean or hose off the bed. Do not immerse any part of the bed. The internal electric parts
may be damaged by exposure to water. Hand wash all surfaces of the bed with warm water and mild
detergent. Dry thoroughly. Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine
Bleach products, typically 5.25% Sodium Hypochlorite in dilutions ranging between 1 part bleach to 100
parts water, and 2 parts bleach to 100 parts water are not considered mild detergents. THESE PROD-
UCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO YOUR BED IF USED IMPROP-
ERLY. If these types of products are used to clean Stryker patient care equipment, measures must be
taken to insure the beds are wiped with clean water and thoroughly dried following cleaning. Failure to
properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing
premature corrosion of critical components. Failure to follow the above directions when using these types
of cleaners may void this product’s warranty.
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate.
(Appropriate disinfectant for nylon Velcro should be determined by the hospital.)
Preventative maintenance should be performed at a minimum of biannually to ensure all bed features
are functioning properly. Close attention should be given to safety features including, but not limited to:
safety side latching mechanisms, frayed electrical cords and components, all electrical controls return
to off or neutral position when released, caster braking systems, no controls or cabling entangled in bed
mechanisms, leakage current 100 MA maximum, scale and bed exit systems calibrated properly.
Always unplug bed during service or cleaning. When working under the bed with the bed in the high posi-
tion, always place blocks under the litter frame and set the brakes to prevent injury in case the Bed Down
switch is accidently pressed.
Explosion Hazard – do not use bed in the presence of flammable anesthetics.
Summary of Contents for 2031
Page 18: ...page 1 13 Bed Symbols ...
Page 27: ...page 3 4 Notes ...
Page 39: ...page 4 12 Notes ...
Page 52: ...page 6 11 Power and Sensor Coil Cord Replacement Illustration A B A B VIEW FROM CENTER OF BED ...
Page 71: ...page 8 6 Notes ...
Page 78: ...Base Assembly and Options page 10 3 FOOT END HEAD END ...
Page 80: ...Base Assembly and Options page 10 5 HEAD END HEAD END ...
Page 81: ...Base Assembly and Options page 10 6 HEAD END LEFT SIDE FOOT END ...
Page 82: ...Base Assembly and Options page 10 7 Typical Both Ends Typical Both Ends ...
Page 101: ...Optional Epic Base Assembly page 10 26 HEAD END HEAD END ...
Page 102: ...Optional Epic Base Assembly page 10 27 HEAD END HEAD END FOOT END ...
Page 103: ...Optional Epic Base Assembly page 10 28 ...
Page 104: ...Optional Epic Base Assembly page 10 29 ...
Page 109: ...Litter Assembly and Options page 10 34 ...
Page 110: ...Litter Assembly and Options page 10 35 GC ...
Page 111: ...Litter Assembly and Options page 10 36 GB ...
Page 112: ...Litter Assembly and Options page 10 37 ...
Page 113: ...Litter Assembly and Options page 10 38 HEAD WALL VIEW ...
Page 114: ...Litter Assembly and Options page 10 39 CPR RELEASE ...
Page 115: ...Litter Assembly and Options page 10 40 CPR SLIDE BRACKET DETAIL ...
Page 116: ...Litter Assembly and Options page 10 41 GROUNDING DETAIL HEAD END FOOT END ...
Page 117: ...Litter Assembly and Options page 10 42 HEAD END FOOT END SIDERAIL EXTENSION CABLE ROUTING ...
Page 123: ...Notes page 10 46 2 ...
Page 127: ...Epic Litter Option Assembly page 10 50 ...
Page 128: ...Epic Litter Option Assembly page 10 51 ...
Page 129: ...Epic Litter Option Assembly page 10 52 ...
Page 132: ...Head End Siderail Assembly Left and Right page 10 55 ...
Page 133: ...Head End Siderail Assembly Left and Right page 10 56 CARRIER ARM ASSEMBLY ...
Page 135: ...Head End Siderail Assembly Left and Right page 10 58 LEFT HEAD END SIDERAIL WIRE ROUTING ...
Page 136: ...Head End Siderail Assembly Left and Right page 10 59 RIGHT HEAD END SIDERAIL WIRE ROUTING ...
Page 147: ...Foot End Siderail Assembly Left and Right page 10 70 ...
Page 177: ...2040 120 4 Optional Pleur Evac Rack with Defibrillator Tray page 10 100 ...
Page 183: ...2025 40 Optional Bed Extender Pad page 10 106 ...