page 7–3
Scale System Diagnostics and Calibration
Verifying Scale Accuracy:
1. Zero the empty bed. Place a known weight on the center of the bed; the heavier the better and no less
than 100 pounds. The displayed weight should be within 1% of the actual weight.
2. If the displayed weight is not accurate, remove the weight from the bed and proceed to the Scale Calibration
section.
Scale Calibration:
NOTE
It requires two people to enable the calibration mode for the scale system.
Raise the siderails when calibrating the scale to avoid getting inaccurate scale readings due to possible inter-
ference between the siderails and the casters.
Calibrate the scale system with a known 200 pound weight. If exactly 200 pounds is not available, the factory
default for calibration will have to be changed as described in step 8.
1.
To enter the calibration mode, unplug the bed’s power cord from the wall socket.
2.
Press and hold down the LBS/KGS button.
3.
While still holding the LBS/KGS button, plug the bed’s power cord into the wall socket.
4.
After two seconds, release the LBS/KGS button. The LCD should read “CALIBRATE SCALE”. The cal-
ibration mode is now active.
5.
Using the up or down arrow button (ZERO or SCALE ON/OFF), toggle through the menu until “INIT. TO
DEFAULTS” is displayed. Press and hold the ENTER button (LBS/KGS) until “SAVING DEFAULTS” is
displayed. Release the ENTER button.
6.
Using the up or down arrow buttons (ZERO or SCALE ON/OFF), toggle through the menu until “CALI-
BRATE SCALE” is displayed.
7.
Press and hold the ENTER button (LBS/KGS). Zero the bed, following the displayed instructions. When
the bed is zeroed, the LCD should display “REF X100=<2 0000”. This is the factory default for 200
pounds. If 200 pounds will be used to calibrate the scale, proceed to step 9.
8.
If exactly 200 pounds is not available, change the display to match the weight you are using. Pressing
the TREND/FOWLER ANGLE button will move the cursor position to the right. Pressing the up arrow
(ZERO) button will increase the numbers. Pressing the down arrow (SCALE ON/OFF) button will de-
crease the numbers. Scroll through the numbers until they match the weight you will use for calibration.
9.
Press and release the SCALE ON/OFF button and the LCD will display “PRESS REV. TREND”. Press
the button with the Reverse Trendelenburg symbol (feet down/head up) until the bed reaches full reverse
Trend. Release the button and the LCD will display “DO NOT TOUCH BED”.
10. The LCD will display “PRESS TREND.” Press the button with the Trendelenburg symbol (feet up/head
down) until the bed reaches full trend. Release the button and the LCD will display “DO NOT TOUCH BED”.
11. The LCD will display “CALIBRATE SCALE”. This indicates the calibration procedure is complete.
12. Exit scale calibration by pressing the up arrow button (ZERO) until the LCD displays “EXIT DIAGNOS-
TICS”. Press the ENTER (LBS/KGS) button to exit.
13. Level the bed at a full up or full down position. Remove the weight and zero the bed.
14. Verify scale accuracy and functionality before returning the bed to service.
Summary of Contents for 2031
Page 18: ...page 1 13 Bed Symbols ...
Page 27: ...page 3 4 Notes ...
Page 39: ...page 4 12 Notes ...
Page 52: ...page 6 11 Power and Sensor Coil Cord Replacement Illustration A B A B VIEW FROM CENTER OF BED ...
Page 71: ...page 8 6 Notes ...
Page 78: ...Base Assembly and Options page 10 3 FOOT END HEAD END ...
Page 80: ...Base Assembly and Options page 10 5 HEAD END HEAD END ...
Page 81: ...Base Assembly and Options page 10 6 HEAD END LEFT SIDE FOOT END ...
Page 82: ...Base Assembly and Options page 10 7 Typical Both Ends Typical Both Ends ...
Page 101: ...Optional Epic Base Assembly page 10 26 HEAD END HEAD END ...
Page 102: ...Optional Epic Base Assembly page 10 27 HEAD END HEAD END FOOT END ...
Page 103: ...Optional Epic Base Assembly page 10 28 ...
Page 104: ...Optional Epic Base Assembly page 10 29 ...
Page 109: ...Litter Assembly and Options page 10 34 ...
Page 110: ...Litter Assembly and Options page 10 35 GC ...
Page 111: ...Litter Assembly and Options page 10 36 GB ...
Page 112: ...Litter Assembly and Options page 10 37 ...
Page 113: ...Litter Assembly and Options page 10 38 HEAD WALL VIEW ...
Page 114: ...Litter Assembly and Options page 10 39 CPR RELEASE ...
Page 115: ...Litter Assembly and Options page 10 40 CPR SLIDE BRACKET DETAIL ...
Page 116: ...Litter Assembly and Options page 10 41 GROUNDING DETAIL HEAD END FOOT END ...
Page 117: ...Litter Assembly and Options page 10 42 HEAD END FOOT END SIDERAIL EXTENSION CABLE ROUTING ...
Page 123: ...Notes page 10 46 2 ...
Page 127: ...Epic Litter Option Assembly page 10 50 ...
Page 128: ...Epic Litter Option Assembly page 10 51 ...
Page 129: ...Epic Litter Option Assembly page 10 52 ...
Page 132: ...Head End Siderail Assembly Left and Right page 10 55 ...
Page 133: ...Head End Siderail Assembly Left and Right page 10 56 CARRIER ARM ASSEMBLY ...
Page 135: ...Head End Siderail Assembly Left and Right page 10 58 LEFT HEAD END SIDERAIL WIRE ROUTING ...
Page 136: ...Head End Siderail Assembly Left and Right page 10 59 RIGHT HEAD END SIDERAIL WIRE ROUTING ...
Page 147: ...Foot End Siderail Assembly Left and Right page 10 70 ...
Page 177: ...2040 120 4 Optional Pleur Evac Rack with Defibrillator Tray page 10 100 ...
Page 183: ...2025 40 Optional Bed Extender Pad page 10 106 ...