ProntoPress-20
Instruction Manual
48
Replace the cooling water in the Recirculation Cooling Unit at
least once a month.
Place a disposable plastic insert (TREPO) in the tank and fold
it over the edge.
Fill the tank with 30 l of water. NB! Too high a level of coolant
in the tank might damage the pump.
Add Struers Additive according to the instructions.
Put the lid on the tank.
Place the pump in the lid
The cooling unit should be checked for cooling water every week
and refilled when the flushing pump cannot reach the cooling
water. Remember to add Struers Additive for Cooling Fluid: One
part of Additive for 33 parts of water.
The concentration of additive should always lie between 2.7 and
3.3%; the pH should be around 9. Replace the cooling water if pH
is
T
8.0.
To check the concentration of additive, use a refractometer (0-10
Brix, see diagram.). Concentration = 2 x Brix. Add Struers
Additive for Cooling Fluid if the concentration is lower than 2.7%.
Add water if the concentration is higher than 3.3%.
Important
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive each time you refill with water.
Checking the
Recirculation Cooling Unit
Refilling the Recirculation
Cooling Unit
Summary of Contents for ProntoPress-10
Page 1: ...Manual No 15077003 Date of Release 11 01 2008 ProntoPress 20 Instruction Manual ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 60: ...ProntoPress 20 Gebrauchsanweisung Handbuch Nr 15077003 Auslieferungsdatum 11 01 2008 ...
Page 61: ......
Page 63: ......
Page 65: ......
Page 121: ...ProntoPress 20 Mode d emploi Mode d emploi no 15077003 Date de parution 11 01 2008 ...
Page 122: ......
Page 124: ......
Page 126: ......
Page 187: ......
Page 188: ......
Page 189: ......
Page 190: ......
Page 191: ......
Page 193: ......